Question 1

When using water traps to help minimize the problems caused by condensation in a heated humidifier ventilator circuit, where should the RCP place the traps?

  1. Near the humidifier
  2. At low points in the circuit
  3. At high points in the circuit
  4. Near patient’s airway

 

Question 2

Which of the following describes the correct position of a properly inserted oropharyngeal airway?

  1. Distal tip a the level of the uvula, flange extending outside the teeth
  2. Distal below the epiglottis, flange extending outside the teeth
  3. Distal tip a the base of the tongue, flange just outside the teeth
  4. Distal tip at the base of the tongue, flange inside teeth

 

Question 3

Which of the following devices can deliver an FiO2 higher than expected if the patient’s tidal volume decreases?

  1. Venturi mask
  2. Simple O2 mask
  3. Nasal cannula
  4. Partial rebreathing mask
  5. Aerosol face mask

 

Question 4

Which of the following devices is appropriate to use for temporarily administering ventilation during CPR when a manual resuscitator is not available?

  1. Guedel airway
  2. Mouth-to-valve mask
  3. Partial rebreathing mask
  4. Berman airway

 

Question 5

Which of the following features incorporated into most endotracheal tubes assist the RCP in verifying proper tube placement?

  1. Length markings on the curved body of the tube
  2. Radiopaque indicator near the tube tip
  3. Murphy eye near the tube tip

 

Question 6

Which of the following humidifiers would be appropriate to use with adult mechanical ventilators?

  1. Bubble
  2. Bubble-jet
  3. HME
  4. Wick

 

Question 7

Which of the following is NOT a consideration when administering ventilation to an infant with a manual resuscitator?

  1. Attaching the resuscitator to an oxygen source
  2. Choosing the proper size mask for the infant
  3. Using a manual resuscitator without a pressure release valve
  4. Choosing the proper size bag for the infant

 

Question 8

Which of the following should the RCP suggest to help minimize the problems caused by condensation in heated humidifier circuits?

  1. Installing water traps in the circuit
  2. Using a heated wire circuit
  3. Setting the heater temperature to 25 degrees C

 

Question 9

Which of the following statements is false about low-flow O2 delivery systems?

  1. The greater the patient’s inspiratory flow, the greater the FiO2.
  2. The patient’s flow usually exceeds that from a low-flow device.
  3. The O2 provided by a low-flow device is diluted with air.
  4. All low-flow devise provide variable O2 concentrations.

 

Question 10

Which of the following types of humidifiers should the RCP select to reduce the accumulation of water in the circuit of a mechanical ventilator?

  1. Heat and moisture exchanger
  2. Heated pass-over humidifier
  3. Heated cascade humidifier
  4. Heated wick humidifier

 

Question 11

Which of the following would be an absolute contraindication to CPAP therapy?

  1. Patient who is tachycardic
  2. Patient with an elevated blood pressure
  3. Patient who is hypoventilating
  4. Patient in acute congestive heart failure

 

Question 12

While an oropharyngeal airway is being inserted into an apparently unconscious patient, the patient suddenly begins to cough violently. What should the RCP do to ensure a patent airway for this patient?

  1. Continue to insert the airway because this response is normal
  2. Spray the back of the throat with lidocaine and reinsert the airway
  3. Withdraw the oropharyngeal airway and insert a nasopharyngeal airway
  4. Perform an emergency tracheostomy

 

Question 13

While analyzing the FiO2 in different areas of a mist tent providing aerosol therapy to a 15-month-old, the RCP notices that the analyzer is fluctuating from 25% to 35%. What action should the RCP take at this time?

  1. Ensure that all of the possible sources of leaks are sealed
  2. Use an air-oxygen blender to provide the FiO2
  3. Use an oxyhood instead of a mist tent
  4. Increase the oxygen flow rate on the flow meter

 

Question 14

While checking the operation of a mist tent being used by a two-year-old child, the CRT notices the absence of mist entering the enclosure. What corrective actions should be taken in this situation?

  1. Tuck the canopy sides under the mattress
  2. Ensure that the canopy zipper is closed
  3. Check the water level in the nebulizer
  4. Clean the nebulizer’s jet

 

Question 15

While performing rounds, a RCP is told by a patient wearing a cannula at 4 L/min O2, “It feels like nothing is coming out.” Which of the following should be examined by the RCP?

  1. Ensure that the cannula is connected to the flow meter
  2. Ensure that the cannula or any of its tubing is not kinked
  3. Ensure that the flow meter is working correctly
  4. Ensure that the humidifier is not leaking

 

Question 16

While using a heated wick humidifier, the physician asks you to ensure that the patient does not receive a burn to the airway. To monitor this situation, you should place a thermistor probe or thermometer:

  1. In the inspiratory limb of the circuit, near the paitent wye
  2. In the expiratory limb of the circuit, near the wye
  3. In the expiratory limb of the circuit, just after the heater-tubing interface
  4. In the inspiratory limb of the circuit, just after the drainage bag

 

Question 17

While using a manual resuscitator during a resuscitation effort, the patient remains cyanotic. What steps should the RCP take to ensure that the highest possible oxygen concentration is being administered?

  1. Use the highest oxygen input flow
  2. Use the longest possible refill time
  3. Connect an oxygen reservoir to the bag

 

Question 18

You note that the air intake port of a venturi mask set to deliver 35% is partially obstructed by the patient’s bedding. Which of the following would you expect to occur?

  1. A decrease in the device’s output flow
  2. An increase in the O2 percent delivered by device
  3. A change in the FiO2 received by the patient


Question 19

Which of the following accessory devices would be appropriate to use with MDI for administration of metaproterenol to a patient who has hand-breath coordination difficulty?

  1. Nose clips
  2. Spacer
  3. Mask
  4. One-way valve

 

Question 20

When using an MDI without a holding chamber, the patient should be instructed to fire the device at what point?

  1. Immediately after a slow inspiration
  2. Immediately as you begin a slow inspiration
  3. Immediately before beginning a slow exhalation
  4. Immediately after beginning a slow exhalation


Question 21

To ensure delivery of the proper drug dosage with an MDI, which of the following must be done before it is used?

  1. The canister valve stem should be cleaned with a pin
  2. The canister should be warmed to hand temperature
  3. The canister should be shaken

 

Question 22

A patient’s prescription for an MDI bronchodilator specifies “2 puffs QID”. What time period should elapse between each puff of a single treatment?

  1. 1 minute
  2. 5 minutes
  3. 30 seconds
  4. 10 seconds

 

Question 23

For which of the following patients would you recommend against using a flow-triggered MDI for a bronchodilator delivery system?

  1. A patient who cannot coordinate MDI firing with inhalation
  2. A patient who has developed acute severe bronchospasm
  3. A teenager with asthma who refuses to use a holding chamber
  4. A stable, elderly patient on maintenance bronchodilator therapy

 

Question 24

Proper use of a DPI requires that the patient be able to do which of the following?

  1. Coordinate firing of the DPI with inspiration
  2. Generate inspiratory flows of 40 L/minute or higher
  3. Exhale forcibly through the device before drug therapy
  4. Inhale slowly and perform a breath hold

 

Question 25

Which of the following patient groups would use of a DPI for bronchodilator administration NOT be recommended?

  1. Infants and children less than 6 years of age
  2. Patients suffering an acute bronchospasm
  3. Patients requiring maintenance therapy

 

Question 26

All of the following are critical components of a CPAP system EXCEPT:

  1. Source gas
  2. Pressure resistor
  3. Oxygen analyzer
  4. A nebulizer

 

Question 27

A patient receiving 60% FiO2 via a t-piece at 12 LPM is receiving a total flow of:

  1. 24 Liters
  2. 36 Liters
  3. 48 Liters
  4. 12 Liters

 

Question 28

An 18-month-old cystic fibrosis patient is hospitalized for secretion problems. Which of the following devices would be best suited for this patient?

  1. Mist tent
  2. Aerosol face mask
  3. Incubator
  4. Oxyhood

 

Question 29

An adult patient has a Shiley cuffed tracheostomy tube inserted and is receiving aerosol therapy form a T-piece. The patient complains of difficulty breathing and the RCP is unable to pass a 14-French suction catheter into the patient’s trachea. Which of the following actions is most appropriate at this time?

  1. Remove patient from T-Piece and begin manual ventilation with a manual resuscitator
  2. Increase the FiO2
  3. Inspect the inner cannula for a mucus plug
  4. Replace the trahceostomy tube

 

Question 30

An alert patient who has a pH of 7.36, PaCO2 of 42 mmHg, and a PaO2 of 40 mmHg while receiving a FiO2 of .60 is considered to be in acute oxygenation failure. The patient will require oxygenation for the next 24 to 72 hours. Which of the following devices would be most appropriate for administering oxygen to his patient?

  1. Nonrebreathing mask
  2. Noninvasive mask CPAP
  3. Mechanical ventilator
  4. Air-entrainment mask

 

Question 31

An infant requires both a precise high O2 and needs to maintain a neutral thermal environment to prevent cold stress. Which of the following systems would best achieve this?

  1. Oxyhood or warmed O2 with blender
  2. Heated incubator
  3. Heated incubator in combination with an oxyhood and blending system
  4. Air-entrainment nebulizer infant mask with added heater

 

Question 32

An intubated patient with a minute ventilation of 12 L/min is connected to a T-piece and air entrainment nebulizer set at 70% O2 with the O2 flowmeter set at 12 L/min. Which of the following is true?

  1. The patient is not receiving 70% O2 throughout inspiration
  2. The set up is not operating as a high flow system for this patient
  3. The O2 flowmeter setting needs to be increased

 

Question 33

At high flow rates, some humidifiers increase the risk of contamination due to the production of:

  1. Low pressures
  2. Microorganisms
  3. Additional heat
  4. Aerosol particles

 

Question 34

During administration of a noninvasive CPAP via mask to a patient with atelectasis, you find it difficult to maintain the prescribed airway pressure. Which of the following are most likely the problems?

  1. System or mask leaks
  2. Outflow obstruction
  3. Improper fit of mask for patient

 

Question 35

How often should HMEs be replaced?

  1. When ventilator circuits are changed out
  2. When condensate is visible
  3. At least once a week
  4. When contaminated by secretions

 

Question 36

How would you estimate the appropriate length for a nasophyaryngeal airway?

  1. Just estimate the length based on age
  2. Measure the distance from the earlobe to the Adam’s apple
  3. Measure the distance from the earlobe to the tip of the nose
  4. Subtract twice the diameter of the tube from its length

 

Question 37

Immediately after an oral intubation attempt on a cardiac arrest victim, you hear bilateral breath sounds during ventilation but observe only a slight color change on a disposable CO2 colorimeter attached to the airway. Which of the following actions would be most appropriate to take?

  1. xtubate the patient and try to reintubate nasally
  2. Confirm correct placement by requesting a chest x-ray
  3. Immediately obtain a arterial blood gas and check PaCO2
  4. Continue ventilation and assume tube is correctly placed

 

Question 38

It is suggested that bubble humidifiers are of limited effectiveness at flows above which of the following?

  1. 4 L/min
  2. 10 L/min
  3. 8 L/min
  4. 6 L/min

 

Question 39

The RCP is checking the aerosol output from an air entrainment nebulizer and notices that the aerosol output has decreased. Which of the following situations would cause a decreased aerosol from the nebulizer?

  1. The absence of a reservoir bag
  2. An increase in the FiO2 setting
  3. A low water level in the reservoir
  4. A defective wick

 

Question 40

The RCP is working with a patient who is receiving 35% O2 from a Venturi mask and notices that the patient has thick secretions. Which of the following are possible actions to take that would help this situation?

  1. Add aerosol through the aerosol collar, attached to the air-entrainment port set at 21%
  2. Switch the patient to a air-entrainment nebulizer operating at 60%
  3. Perform endotracheal suctioning
  4. Increasing the flow on the flow meter

 

Question 41

The removable inner cannula commonly found in tracheostomy tubes serves which of the following purposes?

  1. Aids in routine tube cleaning and trach care
  2. Prevents the tube from slipping into the trachea
  3. Provides a patent airway should it become obstructed

 

Question 42

What should be done with the outer cannula of a tracheostomy tube when the inner cannula is being cleaned?

  1. Left in place
  2. Replaced with an endotracheal tube
  3. Removed for cleaning first
  4. Replaced with a tracheostomy button

 

Question 43

What size ET tube would you select to intubate a 1500 gram newborn infant?

  1. 2.5 mm
  2. 3.0 mm
  3. 3.5 mm
  4. 4.0 mm


Question 44

What would be the most appropriate humidification system for a patient with an artificial airway in place?

  1. Cool humidity
  2. Heated aerosol
  3. Cool water vapor
  4. Heated humidity

 

Question 45

When checking an oxygen delivery system that incorporates a bubble humidifier running at 6 L/min, you occlude the delivery tubing and the humidifier pressure relief immediately pops off. What does this indicate?

  1. Normal, leak-free system
  2. Leak in the delivery tubing
  3. Malfunctioning humidifier
  4. Malfunctioning humidifier