Trigger sensitivity is an important mechanical ventilation setting because it determines how easily a spontaneously breathing patient can initiate a ventilator-assisted breath. The ventilator must recognize the beginning of the patient’s inspiratory effort and respond by delivering the selected form of support.
If triggering requires too much effort, the patient may perform unnecessary respiratory muscle work before assistance begins. If the setting is too sensitive, unwanted breaths may occur.
Proper trigger sensitivity therefore helps reduce work of breathing, improve comfort, and promote better patient-ventilator synchrony.
What Is Trigger Sensitivity?
Trigger sensitivity determines how much change in pressure or flow must occur before the ventilator recognizes that the patient is attempting to inhale.
During assisted ventilation, the patient begins inspiration by contracting the respiratory muscles. This effort produces changes within the respiratory system that can be detected by the ventilator. Depending on the ventilator and selected settings, the machine may monitor airway pressure, circuit flow, or both.
Once the required trigger threshold is reached, the ventilator begins the supported breath.
The goal is to make the ventilator responsive enough that genuine inspiratory efforts receive prompt assistance without making the system so sensitive that unrelated pressure or flow changes cause unwanted breaths.
Trigger sensitivity is therefore a balance between two problems:
- A trigger that is not sensitive enough increases patient effort and may cause ineffective triggering.
- A trigger that is too sensitive can cause autocycling or auto-triggering.
Note: The ideal setting allows the patient to trigger the ventilator with minimal unnecessary effort while maintaining stable breath detection.
The Trigger Variable
Mechanical breaths can be described according to the variables that control different phases of the respiratory cycle. The variable responsible for beginning inspiration is known as the trigger variable.
Triggering describes the transition from expiration to inspiration.
Several variables can theoretically trigger inspiration, including:
- Time
- Pressure
- Flow
- Volume
Time, pressure, and flow are especially important in clinical mechanical ventilation. It is important to understand that triggering describes only how inspiration begins. It does not determine how the rest of the breath is delivered.
For example, a patient may trigger a breath by creating a change in pressure, but the ventilator may then deliver that breath using volume control or pressure control. Likewise, the breath may terminate according to time, flow, or volume depending on the selected ventilator mode. Triggering is therefore only one component of the complete mechanical breath.
Time-Triggered Breaths
A time-triggered breath begins after a predetermined interval has elapsed. The respiratory rate determines this interval. If the ventilator rate is set at 12 breaths per minute, one breath is scheduled approximately every 5 seconds.
The patient does not have to initiate a time-triggered breath. If the patient makes no inspiratory effort, the ventilator begins inspiration automatically when the time interval expires.
Time triggering is especially important for patients who are:
- Apneic
- Heavily sedated
- Under general anesthesia
- Experiencing severe respiratory muscle weakness
- Unable to maintain an adequate spontaneous respiratory rate
Note: Time triggering provides a minimum level of ventilatory support in modes such as assist/control ventilation. As spontaneous breathing returns, patient-triggered breaths may begin occurring between or instead of scheduled machine-triggered breaths, depending on the ventilator mode.
Patient-Triggered Breaths
Patient triggering occurs when the ventilator detects a spontaneous inspiratory effort. Instead of waiting for a timer to expire, the ventilator monitors the breathing circuit for evidence that the patient has attempted to inhale.
Two major forms of patient triggering are:
- Pressure triggering
- Flow triggering
Note: Both methods serve the same basic purpose. They allow the patient to request ventilatory assistance by beginning a spontaneous inspiration. The sensitivity setting determines how strong that request must be before the ventilator responds.
Pressure Triggering
Pressure triggering detects a decrease in airway pressure caused by the patient’s inspiratory effort.
Before inspiration begins, airway pressure is at the established baseline. The patient contracts the inspiratory muscles, causing intrathoracic and alveolar pressures to fall. This pressure change is transmitted through the airway and ventilator circuit.
When the airway pressure falls sufficiently below baseline, the ventilator recognizes the change and begins inspiration. The pressure sensitivity setting determines how large the pressure decrease must be.
Typical pressure-trigger values are often around:
−0.5 to −2 cm H₂O below baseline airway pressure
Some descriptions provide a broader acceptable range of approximately −1 to −5 cm H₂O, although settings closer to −1 or −2 cm H₂O generally require less patient effort.
Understanding Negative Pressure Sensitivity Values
Pressure-trigger values can initially be confusing because a number closer to zero represents greater sensitivity.
For example:
- −1 cm H₂O is more sensitive than −2 cm H₂O.
- −2 cm H₂O is more sensitive than −5 cm H₂O.
- −5 cm H₂O requires more patient effort than −1 cm H₂O.
Suppose baseline airway pressure is 0 cm H₂O. With the sensitivity set at −1 cm H₂O, the patient only needs to decrease airway pressure by approximately 1 cm H₂O before triggering assistance.
With the sensitivity set at −5 cm H₂O, the patient must generate a 5 cm H₂O pressure decrease. Therefore, changing pressure sensitivity from −2 to −5 cm H₂O makes the ventilator less sensitive, not more sensitive. The larger negative value creates a greater inspiratory workload.
Why Triggering Work Matters
Mechanical ventilation is intended to reduce the workload placed on the respiratory muscles. If the patient must generate a large negative pressure before every ventilator-assisted breath, part of that benefit is lost. The patient may perform significant muscular work before inspiratory flow even begins.
This becomes especially important in patients who have:
- Respiratory muscle weakness
- Fatigue
- High ventilatory demand
- Severe obstructive lung disease
- Auto-PEEP
- Reduced respiratory drive
Note: Repeated difficult triggering may substantially increase total work of breathing. A patient who appears to be receiving mechanical support may still be performing considerable respiratory work simply to activate the ventilator.
Trigger Sensitivity and Work of Breathing
Trigger sensitivity directly affects the amount of muscular effort required before ventilator assistance begins. During the triggering phase, the patient’s respiratory muscles begin generating negative intrathoracic pressure. However, gas may not immediately begin flowing from the ventilator.
The pressure change must first be transmitted through several components of the patient-ventilator system, including:
- The lungs
- The airways
- The artificial airway
- The ventilator tubing
- The pressure sensor
- The ventilator’s triggering mechanism
During this period, the patient’s muscles continue working. The airway pressure shown on the ventilator may underestimate the amount of negative pressure actually generated within the patient’s thorax.
For example, the ventilator may register only a modest negative pressure change, while the patient’s respiratory muscles have generated considerably more effort internally. This additional work consumes oxygen and energy without contributing much to actual ventilation.
When repeated over many breaths, difficult triggering may contribute to:
- Respiratory muscle fatigue
- Increased oxygen consumption
- Patient discomfort
- Increased respiratory distress
- Poor patient-ventilator synchrony
Note: Appropriate trigger sensitivity helps minimize this unnecessary workload.
Ventilator Response Time
Trigger sensitivity is closely related to ventilator response time. Response time refers to the delay between the patient’s inspiratory effort and the ventilator’s delivery of assistance.
The process involves several steps. The patient initiates inspiration, pressure or flow changes occur in the circuit, the trigger threshold is reached, the ventilator recognizes the signal, and the inspiratory valve begins delivering flow. Any delay during this sequence increases the amount of work performed before assistance begins.
Factors that can influence response time include:
- Trigger sensitivity
- Patient respiratory drive
- Ventilator design
- Inspiratory valve performance
- Circuit characteristics
- Inspiratory flow settings
- Auto-PEEP
- Dynamic hyperinflation
- Tubing compliance
- Mechanical malfunction
A patient with a strong respiratory drive may reach the trigger threshold rapidly. A weak patient may have difficulty reaching a demanding setting.
Increasing sensitivity can reduce the amount of effort required, but sensitivity is only one part of the problem. The ventilator must also deliver adequate flow after the breath is triggered.
Flow Triggering
Flow triggering detects spontaneous inspiration by monitoring changes in airflow through the breathing circuit. Many modern ventilators maintain a continuous baseline flow, sometimes called bias flow, through the circuit.
Before the patient begins to inhale, the amount of flow entering and returning from the circuit remains relatively stable. When the patient initiates inspiration, some of that flow is drawn into the lungs. As a result, less flow returns to the ventilator.
The ventilator detects this difference. When the difference reaches the selected flow-trigger threshold, the ventilator initiates inspiration.
A commonly used starting value for flow triggering is approximately:
1 to 2 L/min
Some systems describe the trigger as a selected amount below the baseline bias flow. For example, the ventilator may recognize inspiration when returned flow decreases by approximately 2 L/min compared with the flow entering the circuit.
Flow Triggering Versus Pressure Triggering
Flow triggering has traditionally been considered easier for some patients because the ventilator can detect inspiratory effort before the patient creates a substantial negative airway pressure. This may reduce triggering work and shorten response time.
Older ventilators often demonstrated lower trigger work with flow triggering compared with pressure triggering. Modern ventilators, however, may have highly responsive pressure-trigger systems that perform similarly.
Therefore, the choice between pressure and flow triggering is often less important than ensuring that the selected system:
- Detects genuine patient effort quickly
- Requires minimal unnecessary work
- Does not produce false triggering
- Works well with the patient’s respiratory mechanics
Note: Some ventilators allow pressure and flow sensitivity to remain active simultaneously. Whichever threshold is reached first can initiate the breath.
Is Flow Triggering Always Better?
Flow triggering is not automatically superior for every patient. Patients with respiratory muscle weakness may find flow triggering easier because they may be unable to generate a large negative pressure.
However, some patients with severe airflow obstruction, such as COPD or asthma, may not necessarily trigger more effectively using flow sensitivity. The best method depends on the patient and ventilator.
The clinician should evaluate actual patient effort, ventilator graphics, comfort, respiratory pattern, and the presence of auto-PEEP rather than assuming that one triggering method is always best.
What Happens When Sensitivity Is Not Sensitive Enough?
A trigger that is insufficiently sensitive requires too much patient effort. Two major problems can result:
Delayed Triggering
The patient initiates inspiration and eventually reaches the trigger threshold, but only after substantial effort. The ventilator breath occurs later than the patient’s intended inspiration.
The patient may generate a visible negative deflection on the pressure waveform before the mechanical breath begins. A large or prolonged negative deflection suggests excessive triggering work.
Ineffective Triggering
The patient attempts to inhale but fails to reach the ventilator’s trigger threshold. No mechanical breath is delivered. This is also called an ineffective trigger or missed trigger.
The patient performs respiratory muscle work without receiving the intended support. If ineffective efforts occur repeatedly, they can significantly increase work of breathing and contribute to fatigue.
Signs of Ineffective Triggering
Clinical findings may include:
- Visible inspiratory efforts without corresponding ventilator breaths
- Accessory muscle use
- Tachypnea
- Nasal flaring
- Agitation
- Diaphoresis
- Tachycardia
- Distressed facial expression
- Chest-abdominal asynchrony
- Negative pressure deflections between ventilator breaths
- Fewer displayed breaths than the patient’s apparent respiratory efforts
Note: Ventilator waveforms can be especially useful because they may reveal small pressure or flow changes caused by efforts that never produce a supported breath.
Correcting Ineffective Triggering
When ineffective triggering occurs, the clinician should not automatically change sensitivity without further assessment.
Possible corrective actions include:
- Increase trigger sensitivity
- Consider flow triggering
- Assess for auto-PEEP
- Evaluate sedation level
- Assess respiratory muscle strength
- Check for airway obstruction
- Clear secretions if present
- Evaluate the artificial airway
- Assess ventilator tubing and circuit function
- Ensure inspiratory flow is adequate
Note: The underlying cause should guide treatment.
What Happens When Sensitivity Is Too Sensitive?
A trigger that is excessively sensitive may cause the ventilator to interpret nonpatient events as spontaneous inspiratory efforts. The result is autocycling, also called auto-triggering.
During autocycling, the ventilator delivers a breath even though the patient did not actually attempt to inhale.
Potential causes include:
- An excessively sensitive trigger threshold
- Circuit leaks
- Endotracheal tube cuff leaks
- Condensate in the ventilator tubing
- Movement
- Cardiac oscillations
- Circuit disturbances
- Malfunctioning components
Note: The ventilator may display a respiratory rate higher than the patient’s true spontaneous rate. This can be mistaken for tachypnea unless the clinician observes the patient carefully.
Correcting Autocycling
Potential interventions include:
- Decreasing trigger sensitivity
- Checking for leaks
- Correcting cuff leaks
- Draining condensate
- Inspecting the breathing circuit
- Replacing filters if necessary
- Evaluating the trigger method
- Considering a change between pressure and flow triggering
Note: The goal is not to make triggering as sensitive as possible. The goal is to make triggering easy while maintaining accurate detection of true inspiratory efforts.
Auto-PEEP and Trigger Sensitivity
Auto-PEEP is one of the most important causes of difficult triggering. Auto-PEEP, also called intrinsic PEEP, occurs when the patient does not completely exhale before the next breath begins.
Gas remains trapped in the lungs, creating positive pressure at end expiration. This pressure creates an additional load that the patient must overcome before the ventilator can detect an inspiratory effort.
For example, suppose:
- Auto-PEEP is 6 cm Hâ‚‚O.
- Pressure sensitivity is set at −2 cm H₂O.
The patient must first generate enough inspiratory pressure to overcome the 6 cm H₂O of trapped positive pressure. Only then can airway pressure fall an additional 2 cm H₂O to reach the ventilator trigger threshold.
The total pressure change required may therefore be approximately:
6 cm Hâ‚‚O + 2 cm Hâ‚‚O = 8 cm Hâ‚‚O
This is far greater than the trigger setting alone suggests.
Why Sensitivity Alone Cannot Correct Auto-PEEP
When auto-PEEP is the major cause of ineffective triggering, simply making the trigger more sensitive may not solve the underlying problem. The patient must still overcome intrinsic positive pressure before a pressure or flow change can be transmitted to the ventilator. Auto-PEEP should therefore be identified and treated directly.
Possible contributing factors include:
- High respiratory rate
- Large tidal volumes
- Short expiratory time
- High minute ventilation
- Airway obstruction
- Bronchospasm
- Secretions
- Mucus plugging
- Reduced elastic recoil
- Inadequate inspiratory flow that prolongs inspiratory time
- Patient-ventilator asynchrony
Recognizing Auto-PEEP
Ventilator graphics can provide important clues. On a flow-time waveform, normal expiratory flow should return to baseline before the next inspiration begins.
If expiratory flow remains below baseline when the next breath starts, exhalation is incomplete. This strongly suggests dynamic hyperinflation and possible auto-PEEP.
Note: Auto-PEEP can also be measured using an end-expiratory pause in an appropriately passive patient.
Reducing Auto-PEEP
Treatment depends on the cause. Potential interventions include:
- Decrease respiratory rate
- Reduce tidal volume when appropriate
- Increase inspiratory flow during volume-controlled ventilation
- Shorten inspiratory time
- Allow more time for expiration
- Treat bronchospasm
- Administer bronchodilator therapy when indicated
- Clear airway secretions
- Correct artificial airway obstruction
In selected patients with small-airway collapse and dynamic hyperinflation, external PEEP may be applied below the intrinsic PEEP level to reduce the inspiratory threshold load.
The clinician must assess the complete clinical situation rather than relying only on the displayed trigger setting.
Trigger Sensitivity in Assist/Control Ventilation
Assist/control ventilation allows both patient-triggered and time-triggered breaths. If the patient generates an inspiratory effort that reaches the trigger threshold, the ventilator delivers an assisted breath.
If the patient does not trigger a breath before the preset interval expires, the ventilator delivers a control breath. The set respiratory rate therefore establishes a minimum rate.
For example, a postoperative patient may initially be apneic because of anesthesia or sedation. The ventilator provides time-triggered breaths.
As the patient awakens and spontaneous respiratory activity returns, the patient begins triggering additional breaths. At this stage, proper sensitivity becomes important. If triggering is too difficult, the patient may perform unnecessary respiratory work. If triggering is too easy, autocycling may occur.
Trigger Sensitivity in SIMV
Sensitivity also plays an important role in synchronized intermittent mandatory ventilation. During SIMV, mandatory breaths can be synchronized with spontaneous patient efforts.
Before the ventilator delivers a scheduled mandatory breath, it monitors for an inspiratory effort during a synchronization window.
If the patient triggers during this interval, the ventilator delivers the mandatory breath in coordination with the spontaneous effort. If no patient effort occurs, the ventilator delivers the mandatory breath according to the time-triggered schedule.
The purpose of synchronization is to avoid delivering a mandatory breath at an inappropriate point in the patient’s spontaneous respiratory cycle. Trigger sensitivity must remain functional for this process to occur properly.
Trigger Sensitivity in Pressure Support Ventilation
Pressure support ventilation depends heavily on patient triggering because the patient typically initiates each supported breath. Once the ventilator recognizes the trigger, it raises airway pressure to the selected pressure-support level.
Because the patient is responsible for initiating the breath, trigger sensitivity has a major effect on comfort and work of breathing. A patient who must struggle to initiate each pressure-supported breath may derive less benefit from the support than expected.
Trigger sensitivity should therefore be evaluated whenever a patient on pressure support appears tachypneic, fatigued, uncomfortable, or poorly synchronized with the ventilator.
Trigger Asynchrony
Patient-ventilator asynchrony occurs when the patient’s respiratory efforts and the ventilator’s actions do not occur in appropriate coordination. Trigger asynchrony specifically involves problems with the beginning of inspiration.
Important forms include:
- Ineffective triggering
- Delayed triggering
- Autocycling
- Double triggering
Note: Not all of these problems are caused directly by the sensitivity setting, but trigger sensitivity is often part of the assessment.
Double Triggering
Double triggering occurs when two ventilator breaths are delivered with little or no expiratory interval between them. The patient’s inspiratory effort continues after the first ventilator breath ends, causing another breath to be triggered almost immediately.
Double triggering is often related to a mismatch between the patient’s neural inspiratory time or ventilatory demand and the breath delivered by the machine. The correction therefore may involve more than trigger sensitivity.
During volume-controlled ventilation, possible adjustments include:
- Increase inspiratory flow
- Increase tidal volume when clinically appropriate
During pressure-controlled ventilation, possible interventions include:
- Increase inspiratory time
- Adjust inspiratory pressure
- Modify rise time
During pressure support ventilation, options may include:
- Increase pressure support
- Adjust the cycling criterion
Note: Double triggering should therefore be interpreted as part of the overall pattern of patient-ventilator interaction.
Trigger Asynchrony Versus Flow Asynchrony
A patient may successfully trigger a ventilator breath but still feel that the ventilator is not providing enough gas. This is a flow problem rather than a trigger problem.
Inadequate Inspiratory Flow
Inadequate flow occurs when the ventilator’s inspiratory flow does not meet the patient’s demand. The patient may continue generating strong inspiratory effort after the breath has already begun. During volume-controlled ventilation, the pressure waveform may develop a concave or scalloped appearance.
Possible interventions include:
- Increase inspiratory flow
- Change the inspiratory flow pattern
- Consider pressure-controlled ventilation
- Evaluate causes of increased ventilatory demand
Excessive Inspiratory Flow
Excessive flow can also produce poor synchrony. During pressure-controlled ventilation, airway pressure may rise too rapidly. During volume-controlled ventilation, excessive flow may produce abnormal waveform patterns and discomfort.
Possible adjustments include:
- Decrease inspiratory flow
- Reduce inspiratory pressure when appropriate
- Lengthen rise time
Note: Sensitivity and inspiratory flow should therefore be evaluated separately.
Trigger Asynchrony Versus Cycling Asynchrony
Cycling determines when inspiration ends and expiration begins. A patient may trigger the ventilator appropriately and receive adequate flow but still become asynchronous if the ventilator ends inspiration too early or too late. This is a cycling problem.
Possible adjustments can involve:
- Inspiratory time
- Inspiratory-to-expiratory ratio
- Flow-cycling criteria
- Pressure-support settings
- Ventilator mode
Note: Understanding whether the problem occurs at the beginning, middle, or end of inspiration helps identify the correct setting to change.
Using Ventilator Graphics to Evaluate Triggering
Ventilator graphics are valuable tools for recognizing trigger problems.
Common scalar waveforms include:
- Pressure versus time
- Flow versus time
- Volume versus time
Note: Each can provide information about patient-ventilator interaction.
Pressure-Time Waveform
A small negative deflection before a patient-triggered breath represents inspiratory effort. A large or prolonged negative deflection may indicate that the patient is working too hard to trigger the ventilator.
A negative deflection that is not followed by a mechanical breath may represent an ineffective trigger.
Flow-Time Waveform
The flow waveform can help identify patient effort and incomplete expiration. If expiratory flow does not return to baseline before the next breath, auto-PEEP should be suspected.
Patient efforts may also create small changes in flow that do not result in ventilator triggering.
Volume-Time Waveform
Volume graphics are useful for evaluating leaks and breath delivery. Leaks can interfere with trigger performance and contribute to autocycling, particularly with highly sensitive flow-trigger settings.
Patient Assessment Is Essential
Ventilator graphics should always be interpreted together with direct observation of the patient. The patient may provide clues that are not obvious from numerical ventilator values.
Signs of poor synchrony include:
- Accessory muscle use
- Nasal flaring
- Paradoxical chest and abdominal movement
- Tachypnea
- Tachycardia
- Anxiety
- Diaphoresis
- Visible inspiratory efforts between machine breaths
- Active exhalation during ventilator inspiration
- Distressed facial expression
Note: An awake patient may also be able to describe difficulty initiating breaths or feeling that the ventilator is not providing enough flow.
Troubleshooting Trigger Sensitivity
When a trigger problem is suspected, assessment should be systematic.
Assess the Patient
Determine whether the patient’s respiratory drive or muscular ability has changed.
Consider:
- Sedation level
- Level of consciousness
- Respiratory muscle weakness
- Fatigue
- Increased ventilatory demand
- Worsening respiratory disease
- Pain
- Anxiety
Assess the Airway
Check for factors that increase resistance or interfere with pressure transmission.
These may include:
- Secretions
- Mucus plugging
- Kinking
- Endotracheal tube obstruction
- Bronchospasm
Assess the Circuit
Inspect the ventilator circuit for:
- Leaks
- Condensate
- Loose connections
- Obstruction
- Faulty filters
- Equipment malfunction
Note: Leaks are particularly important when autocycling is occurring.
Evaluate Auto-PEEP
Look for incomplete exhalation on the flow waveform and assess whether dynamic hyperinflation is increasing the patient’s trigger workload.
Evaluate Trigger Sensitivity
If the patient is making clear inspiratory efforts but the ventilator responds late or not at all, the trigger may need to be made more sensitive. If breaths occur without genuine patient effort, the trigger may need to be made less sensitive.
Evaluate Inspiratory Flow
Do not assume that every sign of respiratory distress is caused by trigger sensitivity. A patient may trigger normally but continue struggling because inspiratory flow is insufficient.
Typical Initial Trigger Settings
Trigger settings vary by ventilator and clinical situation, but common starting values for adults include approximately:
- Pressure triggering: −0.5 to −2 cm H₂O
- Flow triggering: 1 to 2 L/min
Note: These values should be treated as starting points rather than universal settings. The correct trigger depends on the individual patient’s effort and ventilator response. A numerical value is only appropriate if it produces the desired clinical result.
Common Trigger Sensitivity Errors
Several mistakes can occur when adjusting sensitivity.
- Mistaking a More Negative Number for Greater Sensitivity: A pressure trigger of −5 cm H₂O is less sensitive than −1 cm H₂O because the patient must generate a larger pressure decrease.
- Making the Trigger Extremely Sensitive:Â Setting the trigger as close as possible to spontaneous activation may appear beneficial, but excessive sensitivity can cause autocycling.
- Ignoring Auto-PEEP:Â A patient may appear unable to trigger because of intrinsic PEEP rather than because the selected sensitivity is inappropriate.
- Ignoring Flow Demand:Â Triggering may occur normally, but the patient may continue to struggle because the ventilator’s inspiratory flow is inadequate.
- Adjusting the Ventilator Without Examining the Patient:Â Waveforms and ventilator values are important, but visible patient effort remains essential for identifying the actual problem.
Clinical Goals of Trigger Sensitivity
The primary goals of trigger sensitivity adjustment are to:
- Recognize genuine inspiratory efforts promptly
- Reduce unnecessary respiratory muscle work
- Prevent missed or delayed triggers
- Avoid autocycling
- Improve comfort
- Improve patient-ventilator synchrony
- Support effective spontaneous breathing
- Reduce avoidable metabolic workload
Note: Trigger sensitivity should always be considered as part of the complete patient-ventilator system. A patient’s ability to trigger depends not only on the numerical threshold but also on airway resistance, respiratory muscle strength, circuit characteristics, respiratory drive, auto-PEEP, ventilator design, and inspiratory flow.
Trigger Sensitivity Practice Questions
1. What is trigger sensitivity in mechanical ventilation?
Trigger sensitivity is the amount of patient effort required to initiate a ventilator-assisted breath.
2. What is the purpose of the trigger sensitivity setting?
The purpose is to allow the ventilator to detect a genuine inspiratory effort with minimal unnecessary patient work.
3. What is the trigger variable?
The trigger variable is the variable that determines when inspiration begins during mechanical ventilation.
4. Which three trigger variables are most clinically important?
Time, pressure, and flow are the most clinically important trigger variables.
5. What is a time-triggered breath?
A time-triggered breath begins automatically after a preset interval has elapsed.
6. If a ventilator rate is set at 12 breaths/min, approximately how often is a time-triggered breath delivered?
A time-triggered breath is delivered approximately every 5 seconds.
7. What is a patient-triggered breath?
A patient-triggered breath begins when the ventilator detects a spontaneous inspiratory effort from the patient.
8. What are the two common methods of patient triggering?
The two common methods are pressure triggering and flow triggering.
9. What occurs during pressure triggering?
The patient creates a decrease in airway pressure, and the ventilator initiates inspiration when the selected pressure threshold is reached.
10. What is a commonly recommended pressure trigger sensitivity range?
A commonly recommended pressure trigger sensitivity is approximately −1 to −2 cm H₂O below baseline airway pressure.
11. Which pressure trigger setting is more sensitive, −1 cm H₂O or −5 cm H₂O?
A setting of −1 cm H₂O is more sensitive because it requires less patient effort.
12. What happens when pressure sensitivity is changed from −2 to −5 cm H₂O?
The ventilator becomes less sensitive, requiring the patient to generate a greater negative pressure to trigger a breath.
13. How does an insensitive trigger setting affect work of breathing?
An insensitive trigger setting increases work of breathing because the patient must exert more inspiratory effort before ventilator assistance begins.
14. What is ventilator response time?
Ventilator response time is the delay between the patient reaching the trigger threshold and the ventilator beginning inspiratory flow.
15. What can happen when ventilator response time is prolonged?
The patient may continue generating inspiratory effort while waiting for assistance, increasing work of breathing and dyssynchrony.
16. What is flow triggering?
Flow triggering is a method in which the ventilator detects a change in circuit flow caused by the patient’s inspiratory effort.
17. How does the ventilator detect a patient effort during flow triggering?
The ventilator detects a reduction in returning circuit flow as some of the baseline flow is drawn into the patient’s lungs.
18. What is a common initial flow-trigger sensitivity setting?
A common initial flow-trigger sensitivity setting is approximately 1 to 2 L/min.
19. What is one potential advantage of flow triggering?
Flow triggering may reduce the effort required to initiate a breath because the ventilator can detect inspiration before a large pressure decrease develops.
20. What is ineffective triggering?
Ineffective triggering occurs when the patient makes an inspiratory effort but the ventilator fails to deliver a supported breath.
21. What is delayed triggering?
Delayed triggering occurs when the ventilator eventually responds to a patient effort, but only after the patient has performed excessive inspiratory work.
22. What is autocycling?
Autocycling occurs when the ventilator delivers a breath without a genuine inspiratory effort from the patient.
23. What can cause autocycling?
Autocycling can be caused by excessive trigger sensitivity, circuit leaks, condensate, movement, or other disturbances in the ventilator system.
24. How does auto-PEEP make triggering more difficult?
Auto-PEEP creates an additional positive-pressure load that the patient must overcome before reaching the ventilator trigger threshold.
25. A patient has 6 cm H₂O of auto-PEEP and a pressure trigger set at −2 cm H₂O. Approximately how much total pressure change may be required to trigger the ventilator?
The patient may need to generate approximately 8 cm Hâ‚‚O of total pressure change to trigger the ventilator.
26. Why should trigger sensitivity not be set as sensitive as possible?
If trigger sensitivity is excessive, the ventilator may respond to nonpatient disturbances and deliver unwanted breaths.
27. What is the main goal when adjusting trigger sensitivity?
The goal is to minimize the patient’s triggering effort while preventing false triggering.
28. How can respiratory muscle weakness affect triggering?
Respiratory muscle weakness can make it difficult for the patient to generate enough pressure or flow change to reach the trigger threshold.
29. Why might a weak patient benefit from a more sensitive trigger setting?
A more sensitive setting reduces the amount of inspiratory effort needed to initiate ventilator assistance.
30. Why should inspiratory flow be evaluated when a patient appears asynchronous with the ventilator?
A patient may trigger the ventilator normally but still struggle if the delivered inspiratory flow does not meet the patient’s demand.
31. What waveform finding may indicate excessive patient effort before a triggered breath?
A large or prolonged negative deflection on the pressure-time waveform may indicate excessive triggering effort.
32. What waveform finding may suggest an ineffective inspiratory effort?
A pressure or flow deflection that is not followed by a ventilator breath may indicate an ineffective trigger.
33. What is patient-ventilator synchrony?
Patient-ventilator synchrony is proper coordination between the patient’s respiratory effort and the timing and delivery of ventilator assistance.
34. How can poor trigger sensitivity contribute to patient-ventilator dyssynchrony?
Poor trigger sensitivity can delay or prevent ventilator assistance when the patient attempts to inhale.
35. What clinical signs may suggest trigger asynchrony?
Accessory muscle use, tachypnea, agitation, visible inspiratory efforts, and mismatched patient and ventilator breaths may suggest trigger asynchrony.
36. Why can difficult triggering increase oxygen consumption?
The respiratory muscles use additional energy while generating the effort required to activate the ventilator.
37. In assist/control ventilation, what happens when the patient reaches the trigger threshold?
The ventilator delivers an assisted breath.
38. In assist/control ventilation, what happens if the patient does not trigger before the preset interval expires?
The ventilator delivers a time-triggered control breath.
39. What role does the preset respiratory rate serve in assist/control ventilation?
It establishes the minimum number of breaths the ventilator will provide.
40. How is trigger sensitivity used during SIMV?
Trigger sensitivity allows the ventilator to recognize a spontaneous effort and synchronize a mandatory breath with the patient.
41. What happens during the synchronization period in SIMV?
The ventilator monitors for a patient inspiratory effort before delivering the scheduled mandatory breath.
42. What happens if the patient does not trigger during the SIMV synchronization period?
The ventilator delivers the mandatory breath when the full time interval expires.
43. Why is trigger sensitivity especially important during pressure support ventilation?
The patient typically initiates each pressure-supported breath, so difficult triggering can substantially increase work of breathing.
44. What is dynamic hyperinflation?
Dynamic hyperinflation is progressive air trapping that occurs when the lungs do not fully empty before the next breath begins.
45. What ventilator waveform finding suggests auto-PEEP?
Expiratory flow that fails to return to the zero baseline before the next inspiration suggests auto-PEEP.
46. How can increasing inspiratory flow help reduce auto-PEEP during volume-controlled ventilation?
Increasing inspiratory flow shortens inspiratory time and allows more time for expiration.
47. How can decreasing the respiratory rate help reduce auto-PEEP?
A lower respiratory rate increases the amount of time available for complete exhalation.
48. Why can bronchospasm contribute to difficult triggering?
Bronchospasm increases airway resistance and promotes air trapping, which can increase auto-PEEP and the effort needed to trigger the ventilator.
49. How can circuit leaks affect trigger sensitivity?
Circuit leaks can create pressure or flow changes that the ventilator may incorrectly interpret as patient inspiratory efforts.
50. Why should the clinician assess the patient before changing trigger sensitivity?
Triggering problems may result from respiratory drive, muscle weakness, airway obstruction, auto-PEEP, sedation, or other factors rather than the trigger setting alone.
51. Why can excessive sedation contribute to ineffective triggering?
Excessive sedation can reduce respiratory drive, making it harder for the patient to generate enough effort to reach the trigger threshold.
52. How can an obstructed artificial airway affect triggering?
An obstructed artificial airway increases resistance and can make it more difficult for the patient to generate the pressure or flow change needed to trigger a breath.
53. Why should condensate be removed from the ventilator circuit?
Condensate can disturb pressure or flow measurements and contribute to inappropriate triggering or autocycling.
54. What is the relationship between trigger sensitivity and respiratory drive?
Patients with stronger respiratory drive may reach the trigger threshold more easily, while patients with weak drive may require a more sensitive setting.
55. Can trigger sensitivity alone determine whether a breath is pressure controlled or volume controlled?
No. Trigger sensitivity determines how inspiration begins, not how the breath is controlled after triggering occurs.
56. What is the main difference between pressure triggering and flow triggering?
Pressure triggering detects a decrease in airway pressure, while flow triggering detects a change in circuit flow.
57. Why can the pressure measured at the ventilator underestimate actual patient effort?
The patient may generate greater negative intrathoracic pressure before that pressure change is fully transmitted through the airway and circuit to the ventilator sensor.
58. What is the metabolic consequence of excessive trigger work?
Excessive trigger work increases energy expenditure and oxygen consumption by the respiratory muscles.
59. Why can trigger asynchrony interfere with ventilator liberation?
Repeated ineffective or delayed triggering can increase respiratory muscle workload and make spontaneous breathing more difficult to tolerate.
60. What is a potential benefit of external PEEP in selected patients with auto-PEEP?
External PEEP may reduce the inspiratory threshold load required to trigger the ventilator when applied appropriately below intrinsic PEEP.
61. Why should external PEEP be applied cautiously in patients with auto-PEEP?
Excessive external PEEP can worsen hyperinflation, so it should be adjusted carefully according to the patient’s condition.
62. What is double triggering?
Double triggering occurs when two ventilator breaths are delivered in rapid succession with little or no expiratory time between them.
63. What does double triggering often indicate?
Double triggering often indicates that the ventilator breath does not adequately match the patient’s inspiratory demand or neural inspiratory time.
64. How may increasing inspiratory flow help correct double triggering during volume-controlled ventilation?
Increasing inspiratory flow can shorten inspiratory time and better match the patient’s inspiratory demand.
65. How may increasing inspiratory time help correct double triggering during pressure-controlled ventilation?
A longer inspiratory time may allow the ventilator breath to better match the duration of the patient’s inspiratory effort.
66. What is flow asynchrony?
Flow asynchrony occurs when the inspiratory flow delivered by the ventilator does not match the patient’s flow requirement.
67. What waveform pattern may suggest inadequate inspiratory flow during volume-controlled ventilation?
A concave or scalloped pressure waveform during inspiration may suggest inadequate inspiratory flow.
68. How can inadequate inspiratory flow affect the patient?
Inadequate inspiratory flow can increase respiratory effort, discomfort, and patient-ventilator dyssynchrony.
69. What is cycling asynchrony?
Cycling asynchrony occurs when the ventilator ends inspiration at a time that does not match the patient’s natural transition to expiration.
70. Why should trigger asynchrony be distinguished from cycling asynchrony?
The two problems occur at different phases of the breath and require different ventilator adjustments.
71. Which ventilator graphic is especially useful for detecting incomplete exhalation?
The flow-time waveform is especially useful because expiratory flow should return to baseline before the next breath begins.
72. How can volume-time graphics help when troubleshooting triggering problems?
Volume-time graphics can help identify leaks that may interfere with normal trigger detection.
73. Why is direct observation of the patient important when evaluating autocycling?
The clinician must determine whether displayed ventilator breaths are actually associated with visible patient inspiratory efforts.
74. What should be suspected when the ventilator displays more breaths than the patient appears to initiate?
Autocycling or another source of false triggering should be suspected.
75. Why should trigger sensitivity be reassessed as a patient’s clinical condition changes?
Changes in respiratory drive, muscle strength, airway resistance, sedation, and auto-PEEP can alter how easily the patient triggers the ventilator.
76. Why can high circuit compliance contribute to delayed triggering?
High circuit compliance can delay transmission of the patient’s pressure change to the ventilator sensor, increasing response time.
77. How can high circuit dead space affect ventilator response?
High circuit dead space can contribute to a slower ventilator response and increase the effort required before assistance begins.
78. Why can high bias flow contribute to delayed triggering in some ventilator systems?
High bias flow can alter how the ventilator detects changes in circuit flow and may prolong the response to patient effort.
79. What should be considered if trigger sensitivity appears appropriate but the patient still struggles to initiate breaths?
The clinician should evaluate auto-PEEP, airway resistance, respiratory drive, circuit characteristics, and ventilator response time.
80. Why is a pressure trigger setting of 0 cm Hâ‚‚O generally undesirable?
A setting of 0 cm Hâ‚‚O may make the ventilator excessively sensitive and increase the risk of self-cycling.
81. What is self-cycling in mechanical ventilation?
Self-cycling is the delivery of ventilator breaths without an appropriate patient inspiratory effort.
82. How can pressure-volume loops assist with patient-ventilator assessment?
Pressure-volume loops can provide information about work of breathing, compliance, resistance, and possible asynchrony.
83. What does a wider pressure-volume loop generally suggest?
A wider pressure-volume loop generally suggests increased airway resistance.
84. What does a steeper pressure-volume loop generally indicate?
A steeper pressure-volume loop generally indicates greater respiratory system compliance.
85. Why is sensitivity still important when both pressure and flow triggering are available?
The thresholds must still be adjusted so the ventilator responds promptly to true patient efforts without false triggering.
86. What determines which trigger activates the ventilator when pressure and flow triggering are both active?
Whichever trigger threshold is reached first initiates the ventilator breath.
87. Why can severe airflow obstruction increase the difficulty of patient triggering?
Severe airflow obstruction can prolong exhalation, promote air trapping, and increase the inspiratory workload needed to overcome auto-PEEP.
88. Why should secretions be considered when troubleshooting ineffective triggering?
Secretions can increase airway resistance and contribute to difficulty generating the pressure or flow change needed to trigger the ventilator.
89. What is the purpose of an end-expiratory pause when evaluating auto-PEEP?
An end-expiratory pause can be used to measure intrinsic end-expiratory pressure in an appropriately passive patient.
90. How can decreasing tidal volume help reduce auto-PEEP?
Decreasing tidal volume can reduce the amount of gas that must be exhaled and help provide more complete lung emptying before the next breath.
91. How can shortening inspiratory time help reduce dynamic hyperinflation?
Shortening inspiratory time allows more time for expiration and may reduce air trapping.
92. Why can a high respiratory rate worsen trigger difficulty in a patient with obstructive disease?
A high respiratory rate shortens expiratory time, which can increase air trapping and auto-PEEP.
93. Why should mechanical malfunction be considered when ventilator triggering becomes abnormal?
A malfunction can interfere with accurate detection of patient effort or delay the delivery of inspiratory flow.
94. What is one reason trigger sensitivity may need adjustment as anesthesia wears off?
As spontaneous respiratory effort returns, the patient may begin attempting to trigger assisted breaths and may require an appropriate sensitivity setting.
95. Why can abdominal-rib cage paradox contribute to delayed ventilator response?
Abnormal chest and abdominal movement can reflect inefficient inspiratory effort and make effective triggering more difficult.
96. What is the relationship between trigger sensitivity and myocardial work?
Poor synchrony and increased work of breathing can raise oxygen demand and may increase myocardial workload.
97. Why can trigger asynchrony increase minute ventilation requirements?
The patient may perform ineffective or poorly supported respiratory efforts, increasing overall ventilatory demand.
98. What should be evaluated when a patient actively exhales while the ventilator is still delivering inspiration?
The clinician should assess for patient-ventilator asynchrony, including problems with inspiratory timing and cycling.
99. Why is trigger sensitivity considered only one part of patient-ventilator interaction?
Successful synchrony also depends on inspiratory flow, cycling, respiratory drive, airway resistance, auto-PEEP, and ventilator mode.
100. What is the overall clinical objective of proper trigger sensitivity adjustment?
The objective is to allow genuine inspiratory efforts to initiate ventilator assistance promptly with minimal unnecessary work while preventing false triggering.
Final Thoughts
Trigger sensitivity determines how readily a mechanical ventilator recognizes a patient’s spontaneous inspiratory effort and begins assisted inspiration. A trigger that requires too much effort can increase work of breathing, delay support, and cause ineffective triggering, while excessive sensitivity can result in autocycling. Pressure and flow triggering can both provide effective support when properly adjusted.
Clinicians should evaluate sensitivity together with inspiratory flow, respiratory drive, ventilator waveforms, circuit integrity, and auto-PEEP. The goal is to provide prompt ventilator assistance with minimal unnecessary patient effort while avoiding breaths that are not caused by genuine inspiration.
Written by:
John Landry is a registered respiratory therapist from Memphis, TN, and has a bachelor's degree in kinesiology. He enjoys using evidence-based research to help others breathe easier and live a healthier life.
References
- Radke OC, Schneider T, Vogel E, Koch T. Effect of Trigger Sensitivity on Redistribution of Ventilation During Pressure Support Ventilation Detected by Electrical Impedance Tomography. Anesth Pain Med. 2015.
